<link rel="stylesheet" href="styles.f3b1fba60ec7970c.css">

Farklı yöntemlerle grid performanslarının ölçümü

Loading...
Thumbnail Image

Date

Journal Title

Journal ISSN

Volume Title

Publisher

Nükleer Bilimler Enstitüsü

DOI

Abstract

There is a wide selection of anti-scatter grids in the market in order to make the best balance of patient radiation dose image quality optimization for specific clinical examinations. The use of grids ismore practical for flat panel systems; they may be removed easily or depending upon the clinical procedure different types can be installed.İn this work, we investigated the possible use of flat panel detectors for the performance test of grids according to IEC (International Electrotechnical Commission) procedures. Comparisons heve been made with a similar technique but using a small ion chamber. Transmissions of primary, scatter and total radiations were measured and some other grid parameters such as bucky factory, contrast improvement factory and grid selectivity were calculated.Our results indicated the importance of grid tests since some of the parameters of some grids found quite different than their nominal and expected performances.