Farklı yöntemlerle grid performanslarının ölçümü
Loading...
Files
Date
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Nükleer Bilimler Enstitüsü
DOI
Abstract
There is a wide selection of anti-scatter grids in the market in order to make the best balance of patient radiation dose image quality optimization for specific clinical examinations. The use of grids ismore practical for flat panel systems; they may be removed easily or depending upon the clinical procedure different types can be installed.İn this work, we investigated the possible use of flat panel detectors for the performance test of grids according to IEC (International Electrotechnical Commission) procedures. Comparisons heve been made with a similar technique but using a small ion chamber. Transmissions of primary, scatter and total radiations were measured and some other grid parameters such as bucky factory, contrast improvement factory and grid selectivity were calculated.Our results indicated the importance of grid tests since some of the parameters of some grids found quite different than their nominal and expected performances.
Description
Keywords
Citation
Collections
Endorsement
Review
Supplemented By
Referenced By
Rights and licensing
openAccess
Google Scholar
Scholar'da Ara ↗ Bu yayında DOI yok — Altmetric/Dimensions/PlumX/BIP! rozetleri DOI gerektirir.
