Analytical Formulae to Calculate the Total Efficiency of an Arbitrarily Positioned Point Source by an Elliptical Cylindrical Detector

dc.contributor.authorABBAS, M. I.; HAMMOUD, S.; IBRAHIM, T.; SAKR, M.
dc.date.accessioned2019-04-24T13:23:36Z
dc.date.available2019-04-24T13:23:36Z
dc.description.abstractIn this paper, a direct analytical method is presented for calculating the absolute efficiency of an elliptical cylindrical detector in the case of an arbitrarily positioned (above the major axis a) point source. The absolute efficiency is required to determine the activity of an unknown radioactive source, taking into account the attenuation of the gamma-ray photons. The validity of the derived analytical expressions was successfully confirmed by the comparison with some published data.
dc.identifier.endpage022tr_TR
dc.identifier.issn2147-7736
dc.identifier.startpage012tr_TR
dc.identifier.urihttp://hdl.handle.net/20.500.12575/63378
dc.identifier.urihttps://doi.org/10.1501/nuclear_0000000009
dc.relationAnkara University Institute of Nuclear Sciencestr_TR
dc.relation.journalJournal of Nuclear Sciencestr_TR
dc.titleAnalytical Formulae to Calculate the Total Efficiency of an Arbitrarily Positioned Point Source by an Elliptical Cylindrical Detectortr_TR